Product Description
" XR50RP
"
PN Technical Datasheet
Portable X ray generator
1 Power supply
"(A) single phase 220V,±22V "
"(B)one phase wire;neutral line; safe ground"
"(C)50Hz/60Hz±1Hz;
Power Frequency 50Hz/60Hz±1Hz"
(D)0.5Ω Power resistance less than 0.5Ω
"(E)The power supply capacity is not less than 8kW"
2 Max output power 5.0kW
3 Operating frequency 30kHz wide pulse modulation"
4 Photographic voltage range 40-110kV, step length 1kV"
5 KV accuracy <=±5%
6 mA range 32mA-100mA, (32,40,50,63,80,100)
7 mA accuracy <= ±10%
8 Exposure time range:0.01-6.3s
9 Time arruracy <=±10%+1ms
10 mAs range 0.32-315mAs,press R10'
11 Max power of x-ray tube:5.3kW
12 X-ray tube anode heat capacity:30kJ
13 X ray tube focus nominal value (mm) 1.8 x 1.8 "
14 Beam Limiter: maximum light field 440mm+10mmX440mm+10mm, light source LED lamp, light source timing 30s, manual knob operation mode"
15 Dimension:297*239*230(mm)
16 Weight:23kg