3 Part Differential Auto Hematology Analyzer AJ-2400
FEATURE
* 23 parameters. Throughput:>60T/H
10.4'' TFT LCD
High accuracy technology for blood processing, only 9.6µL blood sample
Storing 40 thousand samples. Operated by mouse & keyboard
* Advanced Design Excellent Performance
9.6µL venous blood sample, 9.6µL anticoagulant capillary blood, 20µL prediluted capillary blood. One time blood sampling could make two tests under prediluted mode.
Intelligent float discriminator. Manual adjust supported
Dual swirling flow and intelligent fitting technology to ensure reliable PLT counting
Module design, separate hydraulic part & circuit, enhanced anti-interference function
Automatic washing of inside and outside surface of probe
Alarm hint if reagent volume is short
* Simple Operation Low Cost
Automatic dispense, Automatic making dilution inside of instrument, Automatic testing, Automatic cleaning
Windows color operation interface and operated by mouse & keyboard
Adjustable probe. Many kinds of blood sampling container
Automatic sleep and wake up function
Low reagent consumption and low cost
* Convenient Maintenance Safety & Eco-protection
Automatic & Intelligent function of diagnosis and maintenance
Soak, back & forth flush, real time monitoring function of high pressure burning and
automatic block-clearing
Perfect self checking function, Automatic cleaning tubes during power on and power off for reducing the rate of blockage
Fan Shape door for easy maintenance
Eco-friendly design of built-in reagents, saving space & reducing the rate of reagent
contamination
No cyanogen reagents for Eco-protection
* Perfect Quality Traceable Results
Manual and automatic calibration available
Boasts management system for quality control, nine kinds of control files. Mean value, SD value, CV value are provided. Automatic drawing L-J graph
Provide relevant blood control and calibrator
Furnished with original reagents
ISO9001, CMD, CMC and CE approved
* Display all the parameters and three histograms in one screen.
* Several groups of reference values setting
* Automatic storing 40 thousand samples. Combined search mode as per many conditions
* Venous blood 9.6µL, anticoagulant capillary blood 9.6µL, prediluted capillary blood 20µL,convenient and reliable blood sampling
* Eco-protection designed built-in reagent system, saving space and decrease the rate of contamination
TECHNICAL PARAMTERS
* Function: Automated WBC 3-differential, 23 parameters with 3 histograms(WBC, RBC, PLT)
* Principal: Electronic impedance, SFT method for hemoglobin
* Testing Channel: Two channels testing, WBC/HHGB channel and RBC/PLT channel
* Parameters: WBC, LYM#, MID#, GRAN#,LYM%,MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, PCT, RDW-CV, RDW-SD, MPV, PDW, P-LCR and WBC, RBC, PLT three histograms
* Throughput: >60T/H, 24 hours switch on, automatic sleep and wake up function
* Sampling Mode: Optional venous blood, anticoagulant blood and prediluted blood
* Sample Volume: Whole blood ≤9.6µL, capillary blood ≤9.6µL, prediluted capillary blood ≤20µL, the sample volume for repeating two times testing and avoiding second sampling
* Working Mode: Automatic sampling, detecting automatic testing, furnished with dilutor, no need diluting outside. Automatic dispensing lyse and making second diluting automatically
* Block-clearing: Soak and washing back and forth. Real time monitoring function of high pressure burning and automatic block-clearing. Boast the function of monitoring incomplete blockage
* Alarm: Automated system diagnosis, self testing, fault alarm & reagent shortage alarm
* Automated Maintenance: Automated washing during power on and power off, reducing the blockage rate
* Storage: Adopting 2G SD card memory. Automatic storing 40 thousand samples with all parameters and 3 histograms. Combined search mode as per many conditions
* Control Mode: Boasts management system for quality control, nine kinds of control files. Mean value, SD value, CV value are provided. Automatic drawing L-J graph. Provide matched blood control
* Calibration Mode: Automatic calibration and manual calibration
* Reference Value Setting: Man, Woman, Children, Baby and General etc.
* Performance: Repeatability Error Linearity Range
WBC <2.0% 0.0~99.9×109/L
RBC <1.5% 0.00~9.99×1012/L
HGB <1.5% 0~300g/L
MCV <0.4% 40~150fL
HCT <1.0% 0.0~100.0%
PLT <4.0% 0.0~999×109/L
* Carry Over: RBC, WBC, HGB<0.5%, PLT<1.0%
* Operating Interface: Colorful Windows interface, no need connecting computer. Operating by graphic button with mouse and keyboard
* Display: Big size 10.4'' TFT LCD, all parameters and histograms displayed in the same screen
* Print Mode: Internal high-speed thermal printer. Optional connection with specified external printer. English printouts
* Reagents System: Open reagent system. No cyanogen reagents. Provide original reagents.
* Connection Port: USB, RS-232, network interface
REAGENT
Ordinary Package of AJ-2400 (Suggest to send by AIR)
One Set including below:
CODE |
DESCRIPTION |
SPEC |
30201 |
Diluents |
5L |
30202 |
Lyse |
200ml |
30203 |
Detergent |
500ml |
Big Package of AJ-2400 (Suggest to send by SEA)
One Set including below:
CODE |
DESCRIPTION |
SPEC |
30201 |
Diluents |
20L |
30202 |
Lyse |
500ml |
30203 |
Detergent |
1L×5 |