On-line HDF Machine AJ-6008HDF
On-line Haemodiafiltration Machine AJ-6008HDF Funcitons
- On-line HDF
- Self-checking function;
- Carbonate dialysis;
- Double needle dialysis;
- Liquid level detector;
- Bubble detector;
- Blood leakage detector;
- Temperature and electric conductivity monitoring;
- Arterial pressure, venous pressure and transmembrane pressure monitoring;
- Rolling blood pump;
- Heparin pump;
- Dehydration amount controlled by capacity;
- Automatic disinfection cleaning program;
- Stand-by power of blood pump in case of power failure;
- Information display function of display screen.
Technical Parameter
Size &Weight:
Size: 380mmx400x1380mm (L*W*H);
Net Weight approx.: 88KG;
Gross Weight approx.: around 100KG;
Package Size approx.: 650×690×1581mm (L x W x H)
Power supply:
AC220V, 50Hz/60Hz, 10A
Input power: 1500W
Back-up battery: 30 minutes
Working Condition:
Water input pressure: 0.1Mpa~0.6Mpa, 15P.S.I~60P.S.I
Water input temperature: 5ºC~30ºC
Working environment temperature: 10ºC~30ºC at relative humidity <=70%
UF rate:
Flow range: 0ml/h~4000ml/h
Resolution ratio: 1ml
Precision: ±30ml/h
Blood pump & substitution pump:
Blood pump flow range: 10ml/min~600ml/min (diameter: 8mm or 6mm)
Substitution pump flow range: 10ml/min~300ml/min (diameter 8mm or 6mm)
Resolution ratio: 0.1ml
Precision: ±10ml or 10% of reading
Heparin pump:
Syringe size: 20, 30, 50ml
Flow range: 0ml/h~10ml/h
Resolution ratio: 0.1ml
Precision: ±5%
Monitoring system &Alarm setup:
Venous pressure: -180mmHg ~ +600mmHg, ±10mmHg
Arterial pressure: -380mmHg ~ +400mmHg, ±10mmHg
TMP: -180mmHg ~ +600mmHg, ±20mmHg
Dialysate temperature: preset range 34.0ºC~39.0ºC
Dialysate flow: Less than 800 ml/min (Adjustable)
Substitution flow range: 0-28 L/H (on line HDF)
Blood leak detection: Photo chromic alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate.
Bubble detection: Ultrasonic, Alarm when a single air bubble volume is more than 200μl at 200ml/min blood flow
Conductivity: Acoustic-optic
Disinfection/Sanitize:
1. Hot disinfection:
Time: 30 minutes; Temperature: about 80ºC, at flow rate 500ml/min;
- Chemical disinfection:
Time: 30minutes, Temperature: about 36ºC~50ºC, at flow rate 500ml/min;
- Chemical disinfection with heat:
Time: 45minutes, Temperature: about 36ºC~80ºC, at flow rate 50ml/min;
- Rinse:
Time: 10minutes, Temperature: about 37ºC, at flow rate 800ml/min;
Storage Environment:
Storage temperature should be between 5ºC~40ºC, at relative humidity <=80%
Function:
HDF, on-line BPM, Bi-cart and 2 pcs endotoxin filters
Optional function:
Online KT/V, LAN