Customization: | Available |
---|---|
Sterilization: | Sterilization |
Type: | Infusion Pump |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product name | AJ-P300G |
Unique Voice Alarm system | The device triggered the voice prompt and stop injection automatically when there is any malfunction, making the infusion process safer and more reliable. |
Infusion rate | 1ml/h~1,200ml/h |
Purge rate | 100ml/h~1,200ml/h (100 ml/h increment) |
KVO rate | 1ml/h~5 ml/h (1ml/h increment) |
Accuracy | Within ±5% (after correct calibration) |
Infusion Volume | 1ml~9999ml |
Total Infusion Volume | 0.1ml~9999.9ml |
Occlusion detection pressure | High: 800±200 mmHg (106.7±26.7 kPa) Medium: 500±100 mmHg (66.7±13.3 kPa) Low: 300±100 mmHg (40.7±13.3 kPa) |
Bubble detector | Ultrasonic wave detector; detection sensitivity ≥25μL |
Alarms | End of injection, occlusion, door open, bubbles in the tube, wrong setting, low battery, AC power pulled out etc. |
Power | AC100V~240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup |
Operation conditions | Ambient temperature: +5°C~+40°C; Relative humidity: 20~90% Atmospheric pressure: 86.0kpa~106.0kpa |
Storage conditions | Ambient temperature: -30°C~+55°C Relative humidity: ≤95% |
Equipment Classification | Class II, internal power supply, Type BF |
IP grade | IPX1 |
Dimension | 140mm (L) × 157mm (W) × 220mm (H) |
Weight | 1.8kg (net weight) |