Good Price Infant Incubator

Certification: CE
Warm-up Time: ≤ 45min
Power Supply: AC110/220V 60/50Hz
Infant Bed Tilt Angle: +/-10degree
Transport Package: Standard Export Package
Specification: ISO and CE
Customization:
Gold Member Since 2011

Suppliers with verified business licenses

Manufacturer/Factory, Trading Company

Basic Info.

Model NO.
AJ-2301
Trademark
Anjue
Origin
China

Product Description

Brief Introduction

Two temperature control modes: Air mode and Baby mode controlled by micro-computer;
Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately;
Failure alarm indication;
>37C temperature set function is optional;
Disposable skin temperature sensor is optional;
Double wall acrylic glass hood with side door, the bassinet can be pulled out from the side door;
Humidify by natural air flow;
The height of whole unit can be adjusted;
Oxygen supply system;
Observation lamp;
Two types of trolley can be chosen: Common trolley and ambulance type trolley.

Standard Configuration
Main body (including the hood, controller, bassinet, internal battery, and observation lamp), oxygen cylinders, oxygen supply system, skin temperature sensor, I. V. Pole, mattress, trolley.

Optional configuration: >37C override temperature mode, extra internal battery, ambulance type trolley, disposable skin temperature sensor.

Specification

AC Power supply: AC220V-230V/50Hz Or
AC110-120V/50-60Hz

DC Power supply: DC12V/10A or DC24V/6A
Input power: 400VA
Mode of control: Air mode and baby mode controlled by micro-computer
Air temperature control range: 25C-37C(Override mode 37-38C)
Baby temperature control range: 34C-37C(Override mode 37-37.5C)
Accuracy of skin temperature sensor: ≤ 0.3C
Temperature variability: ≤ 1.0C
Temperature uniformity: ≤ 1.5C(mattress is on the horizontal position)
Noise inside hood: ≤ 52dB(A) (Environment noise≤ 45dB(A))
Failure alarm: Over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm and so on.
Operating at temperature equilibrium: 90min(Set Temp 36C--Ambient 15C)

Environmental Temperature
Operating range: 10 C~30C
(Set temperature should be higher 3C than the environment temperature)
Ambient air movement rate: ≤ 1.0m/s

Package

Package: Each unit is packed as two cartons
Size (Main body): 109cm x 65.5cm x 75.5cm
(Cabinet): 138cm x 61cm x 39cm

Gross Weight (Main body): 80Kg
(Cabinet): 53Kg