Feature
* 12.1 inch (resolution: 800*600)high-resolution color folding up touch-screen
* Multi-language support. Chinese, English, Spanish, Polish, Italian, Turkish
* Hand writing support
* ECG Waveforms and auto-diagnosis results preview and editing before printing
* Support ECG work station. PC software can real display ECG waveform
Support PACS system. Doctor can view ECG waveform and diagnosis report from system directly.
* USB socket to connect external Printer. Support A4 paper printing.. Printer brands: LJ2250N, HP P2015D,HP P2055D.
* USB flash disk for data record and recover.
* Recording Calibration system
* Anti-defibrillation and Pacemaker detection
* Auto measurement and auto diagnosis function
* 12-lead simultaneous acquisition, amplification and recording. Digital and full-floating design.
* Various of record system, 210mm, 216mm width thermal printer papers. Correct record ECG waveform, gain, record speed, patient information and measurement report.
* 1:1 ratio for LCD screen waveforms and output waveform.net-grid on screen and record paper are the same. DC and AC power supply, built-in rechargeable Lithium battery, at least 2 hours working time.
* Last 120 seconds 12-leads waveforms reviewable and recordable
* Alphanumeric keyboard, individual capital and numeral key. (optional for touch screen input)
* Sleep/Wake Up. Mode for saving energy and extending LCD life
* 300 ECGs internal memory, Built in 2G SD card to support over 10000 ECG records, the compatible brand: Sandisk, Kingston.
* Support upgrading programming on line per U dish, the brand supported including Kingston, PNY, ADATA, Apace
* Automatic measurement and interpretation tested Verified channel formats: 3×4,3×4+1R,3×4+3R,6×2,6×2+1R,12×1,12×1+T
* 122 kinds of arrhythmia analysis
* Manual/Auto/ Rhythm working modes selectable
* Lead statue map to show ECG leads conductivity
* Three kinds of recording modes: Record and not printout, record and printout, no record only printout
* Sketch of displaying and closing ECG leads. Indication of leads off.
* Using unique high-precision digital filter to eliminate baseline drift and other disturbances, avoid ECG waveform distortion and greatly enhance the anti-baseline drift capability to facilitate waveform detection.
* Long-distance data transferred through network socket
Specification
Lead: Standard 12 leads
Gain:2.5mm/mV,5mm/mV,10mm/mV,AGC(automatic),20mm/mV,10/5mm/mV(classification, front means lead gain, behind means chest lead gain) 20/10mm/mV(classification, front means lead gain, behind means chest lead gain), accuracy ±2%.
Filter
AC Filter::50Hz/60Hz/Close
EMC Filter::25Hz/35Hz/45Hz/Close
Drift Filter:0.05Hz/0.15Hz/0.25Hz/0.50Hz
Lowerpass Filter:70Hz/100Hz/150Hz
Calibration Voltage 1mV ± 1%
Noise Level: < 15µ Vp-p
Frequency Response: 0.05Hz-150Hz
Time Constant: ≥5s
Input Impedance: 50MΩ
Input Circuit Current: < 50nA
Tolerance Voltage: ±650mV
CMRR: ≥105dB
Anti Baseline Drift: Automatic
External input/Output: Input ≥100 KΩ sensitivity 10mm/V ±5%
Output: ≤100Ω, Sensitivity 1V/mV ±5%
Communication Interface: Ethernet, RS232, USB
Recording System: 210mm, 216mm width thermal printer papers
Rhythm lead: single channel and three channels selectable. 12 leads selection for each channel
Recording Mode: Auto/Manual/Rhythm
Recording Format: 12×1+T , 12x1,3x4, 6x2,
3×4+1R,3×4+3R,6×2+1R (R: rhythm lead)
Average Template: 4×3+1R, 6×2+1R recording
Record speed: 5mm/s, 10mm/s, 12.5mm/s, 25mm/s,, 50mm/s
Record Paper: Roller and Z-fold
Roller paper: 210mm, 216mm
z-fold paper: 210mm x 295mm, 210 x 140mm or 216mm x 295mm, 216mm x 140mm
Display: 12.1 inch touch color TFT display (Resolution 800×600)
Safety Standard: I, CF type,
Memory: 300 ECGs, Built in 2G SD card to support over 10000 ECG records
Digital filter Mode
Accessories: 1 x power cable,1 x Patient Cable, 6 x Chest Electrodes, 4 x Limb Electrodes, 1 x Paper Roller, 1 x Record Paper, 1 x Ground Line, 2 x fuses