Anaesthesia Machine
NewBright A4

NewBright A4 Anesthesia Machine, designed with state-of-the-art technologies, is one of high utility with excellent performance and complete functions. Both outstanding and inexpensive, it has the best cost performance that can satisfy any customer.
The Ventilator adopts microcomputer control with high-definition 5.7" LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of Vaporizer from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
Reliable pneumatic transmission system.
Economic and proper configuration of technologies, optional as required by the Customer.
Modularized design and upgrading. Upgrading is available as required by the Customer.
TECHNICAL PARAMETERS
PART OF ANESTHESIA MACHINE
Main Body
High-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of Application:
Adult, Child
Gas Source
O2: 0.28~0.6MPa, N2O: 0.28~0.6MPa
Flow Meter
O2: 0.05~1.0L/min,1.1~10L/min
N2O: 0.05~1.0L/min,1.1~10L/min
O2-N2O Linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.16±0.02Kpa, the flow of nitrous oxide would be cut off.
Flow Rate of Rapid Oxygen Supply
25~75L/min
Low Oxygen Pressure Alarm
There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.
Respiratory Circuit
Working mode: All-close, Semi-close, Semi-open
Releasing pressure: 0.5~7 kPa
Respiratory Bellows
Bellows for adults and children
Tidal volume range: 0~1500 ml
PART OF VENTILATOR
Display Mode
High-definition 5.7" LCD screen display
Ventilation Mode
IPPV, SIPPV, SIMV, PEEP, SIGH, MANUAL
Ventilation Parameters
Tidal volume:50~1500 ml
Rate:2~99 bpm
SIMV rate:2~20 bpm
I:E:2:1~1:8
Inspiratory trigger pressure: -1.0~2.0 kPa
PEEP: 0~2.0 kPa
Pressure range: 1.0~6.0 kPa
SIGH:1.5 times the inspiratory time/ 100 times breath
Parameters For Ventilation Monitoring
Tidal volume, Ventilation volume, IPPV rate, SIMV rate, Total respiratory rate, I/E, Peak pressure of airway, Pressure - time waveform, Flow rate - time waveform, PEEP, Inspiratory trigger pressure
Security Alarm System
Oxygen Concentration Alarm: Upper limit setting range:50%~100%,Low limit setting range:15%~50%
Airway Pressure Alarm: Upper limit setting range: 1.0~6.0 kPa, Low limit setting range: 0.4~2.0 kPa
Per-minute Ventilation Volume Alarm: Upper limit setting range:3~20L/min,Low limit setting range: 0~10L/min
Tidal Volume alarm: Upper limit setting range:100~1500ml,Low limit setting range 0~1400ml
Sustained High Pressure Alarm: It will give alarm when stress have consistently been higher than 25 cmHO for 15s.
Suffocation Alarm: It will give alarm when stress have consistently been higher than 10 cmHO and tidal volume is less than 120ml for 15s.
Tube Escape Alarm: It will give alarm when stress have consistently been less than 2cmHO for 15s.
Power Supply
AC 220V 50Hz