Anesthesia Machine with Ventilator AJ-2102
TECHNICAL PARAMETERS
Part of Anesthesia Machine
*Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
*Scope of application adult, child
*Gas source O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
*Flow meter O2: 0.1~10L/min N2O: 0.1~10L/min
*O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off.
*Flow rate of rapid oxygen supply 25~75L/min
*Low oxygen pressure alarm There will be sound alarm when the oxygen pressure < 0.2MPa
*Vaporizer it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
*Respiratory circuit working mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa
*Respiratory bellows bellows for adults, bellows for children
tidal volume range:0~1,500 ml
Part of Ventilator
*Display mode High-definition 5.7" LCD screen display
*Ventilation mode IPPV,SIPPV,SIMV,PEE,MANUAL,SIGH
*Ventilation parameters
Tidal volume 50~1500ml
Rate 2~99bmp
SIMV rate 2~20bpm
I:E 2:1~1:8
Inspiratory trigger pressure -1.0~2.0 kPa
PEEP 0~2.0 kPa
Pressure range 1.0~6.0 kPa
SIGH 1.5 times the inspiratory time
*Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure
*Monitoring of oxygen concentration (21%~100%)
*Security Alarm System
Oxygen Concentration alarm upper limit setting range (50%~100%)
low limit setting range (15%~50%)
Airway pressure alarm upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there is
no tidal volume input for 15 seconds
Power alarm
*Power AC 220V±10% 50Hz±1Hz
(UPS, Storage battery)
*Items within brackets are optional.