Usage
Orthopedics: Restore bone translocation, reset, fixing
Surgery: Taking foreign bodies out of the body, cardiac catheterization, implantable pacemakers, some of angiography and local photography etc.
Features
1. Continuous pulse fluoroscope, it's convenient to connect digital subtraction system.
2. With mega-pixel digital CCD photography, the image is clearer.
3. With unique image software processing technique, the image is clearer for doctors to operate and diagnose. Standard DICOM interface, it's convenient to exchange information with hospitals.
4. Pulse fluoroscope has the advantage of low dose and clearer image, so it meets the needs of high precision, high difficulty minimally invasive surgery, which well protect the security of operator and patient.
5. Unique double foot brake controller design makes it convenient to control the instrument inside and outside the operation room. This design can also protect the health care professionals by reducing the probability to come into contact with the radiation.
6. Exclusive inside and outside monitor design makes it more convenient to observe the operation process either inside or outside the operation room. At the same time, this design can facilitate the outdoor teaching and the real-time observation of the surgical procedures.
7. With high-quality combined high frequency high-voltage X-ray generator, it can greatly reduce the amount of X-ray irradiation.
8. With a perspective KV, MA automatic tracking feature, so that the image brightness, sharpness automatically in the best condition
9. Adopting imported image intensifier, the quality is stable and reliable, and the image is of good clarity.
10. With unique base electric auxiliary support arm design, so it's safer.
11. A unique hand-held controller design, which can control parameter set, equipment movement and radiation field, so the operation is more convenient.
12. With new design of mainframe, the appearance is compact and beautiful.
Standard Configurations
1. C-arm Mainframe 1 set
2. Combined high-frequency high-voltage X-ray generator and high-frequency power supply (5.0kW, 40 kHz, 120kV) 1 set
3. Imported 9 inch image intensifier 1 set
4. Medical mega-pixel ultra-low illumination digital radiography 1 set
5. Digital acquisition and processing station (See workstation list for detail information) 1 set
6. Imported dense grain grids 1 set
7. Electrical adjustable beam limiting device 1 set
8.19 inch IM medical LCD grayscale display 1 set
9.14 inch high-definition medical VGA interface (no flicker) 1 set
10. Unique hand-held controller 1 set
Category |
Items |
Content |
|
Model |
AJ-112C |
|
Electrical Performance |
High frequency inverter power supply |
Power:3.5KW
Inverter frequency: 40 kHz |
|
|
Automatic fluoroscopy |
Tube voltage :40 kv~120kv
Tube current: 0.3mA~~4mA |
|
|
|
Manual fluoroscopy |
Tube voltage :40 kv~120kv
Tube current: 0.3mA~~4mA |
|
|
|
Pulse fluoroscopy |
Tube voltage:40 kv~120kv, continuous
Tube current : 4.1mA~~30mA continuous |
|
|
|
Photography tube voltage, tube current mAs |
40KV~120KV, 20~100mA , 1~180 mAs |
|
|
X-ray Tube |
X-ray tube special for high frequency |
Large focus:1.5,Small focus :0.3
Thermal capacity: 650kJ (867 kHu) |
|
|
Imaging System |
Image intensifier |
Imported TOSHIBA 9 inch image intensifier |
|
CCD photography |
Mega-Pixel digital medical CCD radiography 12 bit |
|
|
|
|
Monitor |
19 inch IM medical LCD grayscale display
14 inch high-definition medical VGA interface
Center definition is 1000 lines and frequency for picture refresh is 60 Hz. |
|
|
|
CCU (central control) |
Real-time acquisition, continuous adjustable recursion, many images storage, up and down image, left and right image, image patching, LIH (last image freeze) |
|
Structure and performance |
Direction wheel and main wheel |
Direction wheel can rotate in any direction, and main wheel can rotate in ±90°. |
|
|
C-arm |
The up and down electrical stroke of pillar is 400mm.
Forward and backward movement:200mm;
Revolution around horizontal axis : ±180°;
Revolution around vertical axis: ±15°,
Slipping on orbit : 120°(+90°~-30°)
SID: 960 mm;
C-arm open distance: 760mm
C-arm arc depth: 640mm; |
|
|
|
|
Plate holder size |
200mmx250mm(8"x10") or 250mmx300mm(10"x12") |
|