Technical parameters: 3.1 Power: 220V ±22V 50HZ 3.2 Sterilization:high temperature andpressure steam 3.3 Capacity: 16L 3.4 Rated operating temperature: 134 ºC 3.5 Rated operating pressure: 0.21Mpa 3.6 Temperature and pressure control range: 105 ºC ~ 134 ºC / 0.05Mpa ~ 0.22Mpa 3.7 Sterilization time range: 3 ~ 40 minutes adjustable 3.8 liner and sealing the door: SUS304stainless steel liner once entire cast 3.9 Pressure Control: control precision 0.005Mpa 3.10 Safety valve open and close range: 0.23Mpa ~ 0.26Mpa 3.11 Temperature uniformity: ≤ ± 0.5 ºC 3.12 Automatic over-temperature protection device: more than set temperature, the system automatically terminated procedures 3.13 Door interlocked: Only the door closed in place to start the work process, the pressure inside the chamber, the door can not open. 3.14 overpressure safety valve automatically discouraging: more than the set pressure, the pressure release valve to open. 3.15 Input Power: 1800W 3.16 Operating temperature: 0 ºC ~ 40 ºC 3.17 After the whole pulse vacuum drying, the vacuum level: up to-0.8-0.9bar 3.18 Residual temperature of Instrumentsdrying: not more than 0.2% 3.19 Jet steam generator, the chamber temperature uniformity 3.20. It is equipped with B & D test, vacuum test program, and use for steam penetration test. 3.21.Full computer control, automatic operation, automatic fault diagnosis.