AJ-2101A is an economic type of Anesthesia Machine.
1.The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
2.Reliable pneumatic transmission system.
3.Economic and proper configuration of technologies, optional as required by the Customer.
4.Modularized design and upgrading. Upgrading is available as required by the Customer.
TECHNICAL PARAMETERS
1)Part of Anesthesia
Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
Scope of application adult, child
Gas source O2: 0.27~0.55MPa
Flow meter O2: 0.1~1.0L/min 1.1~10L/min
Flow rate of rapid oxygen supply 25~75L/min
Low oxygen pressure alarm There will be sound alarm when the oxygen
pressure < 0.2MPa
Vaporizer It has the function of automatic compensation based on pressure,temperature,and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer.
Respiratory circuit working mode: all-close, semi-close, semi-open
releasing pressure: 0.5~7 kPa
Respiratory bellows bellows for adults, bellows for children
tidal volume range:0~1500 ml
2)Part of Ventilator
Display mode High-definition 4.8" LCD screen display
Ventilation mode IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate)
Ventilation parameters Tidal volume 50~1500ml
Rate 4~80 bpm
I:E 2:1~1:4
Pressure range 1.0~6.0 kPa
Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway, inspiratory trigger pressure, pressure - time waveform
Security Alarm System
Airway pressure alarm upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
TV alarm upper limit setting range 0.5~2.0L
low limit setting range 0.02~1.0L
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there is
no tidal volume input in 15s
Power alarm
Power AC 220V±10% 50Hz±1Hz